ASTM D6319: Nitrile Glove Testing Standards Explained
ASTM D6319 is a product specification for nitrile rubber examination gloves used in medical examinations and diagnostic or therapeutic procedures. It covers areas such as dimensions, freedom from holes, physical properties before and after aging, and powder-related requirements. It does not prove that every nitrile glove is medical grade, nor does a D6319 reference alone establish broad chemical, chemotherapy-drug, viral, sterile, food-contact, or cleanroom suitability. Buyers must connect the current standard to an exact finished SKU and its evidence.
That last step is where many procurement reviews fail. A supplier may show a standard logo, a laboratory may report one test, and a buyer may read an AQL value as though all three mean “certified protection.” They do not. The defensible approach separates the requirement, the method, the sampling plan, and the finished-product record.
What ASTM D6319 covers
The active ASTM record, D6319-19(2023), describes nitrile rubber examination gloves for medical application. Its scope includes ambidextrous or paired products, gloves offered by size, and sterile or nonsterile configurations. The specification addresses finished-glove requirements and test procedures for key properties.
The word “examination” matters. A general industrial nitrile glove does not become a medical examination glove because it uses the same polymer, color, thickness, or factory. Surgical gloves, foodservice gloves, chemical-protective gloves, and cleanroom gloves can fall under different requirements. Intended use and destination market determine the evidence path.
In the United States, FDA’s LZA product classification defines non-powdered nitrile or polymer patient-examination gloves as medical devices and lists recognized consensus standards. The current FDA recognition record for D6319 is partial because powder-related provisions cannot override the US prohibition on powdered medical gloves. Therefore “meets ASTM D6319” is not, by itself, a complete FDA market-access statement.
The four-layer evidence model
Use four layers when evaluating any nitrile glove testing claim.
| Layer | Question | Typical evidence | Common mistake |
|---|---|---|---|
| Product specification | What must this glove type meet? | Current ASTM D6319 edition and market requirements | Treating the standard title as a finished-SKU certificate |
| Test method or practice | How is one property assessed? | D5151 for hole detection, D6124 for residual powder, D6978 for specified chemotherapy-drug permeation | Assuming one method proves every protective claim |
| Sampling and acceptance | How many units are sampled and what result accepts or rejects the lot? | Written sampling plan, inspection level, AQL, sample size, and acceptance/rejection numbers | Reading AQL as the literal defect percentage in every accepted lot |
| Finished-SKU record | Which exact product, sizes, lots, dates, and laboratories produced the result? | Test reports, 510(k) or market file where applicable, declaration/certificate context, labeling, and lot records | Applying family-page data to a different model, size, color, or factory |
The chain is only as strong as its weakest link. A correct method on an unidentifiable sample is not useful supplier evidence. A clear SKU report to an obsolete edition may not satisfy a current contract. A current standard number without the actual result proves very little.
Main D6319 requirement areas
Dimensions and construction
Examination gloves are evaluated for required dimensions and construction characteristics. A buyer should check length, width by size, thickness locations, cuff design, handedness, sterility status, and packaging against the exact specification and purchase agreement.
Thickness alone is not a grade. It affects feel, fit, durability, and sometimes performance, but two gloves with the same nominal thickness can differ in formulation and quality. Ask how thickness is measured, at which locations, on what sample, and with what tolerances.
Freedom from holes
ASTM D6319 incorporates a freedom-from-holes requirement, commonly evaluated using ASTM D5151. The current ASTM D5151 record describes a water-leak method for detecting holes in medical gloves.
D5151 is not a puncture-strength test. It does not tell a user how much force a needle, burr, blade, or wire can withstand. It also does not establish the smallest possible defect detectable under every condition or guarantee that an accepted glove can never fail in use. It is a defined finished-glove hole-detection method within a sampling and acceptance system.
Physical properties before and after aging
Tensile strength and ultimate elongation help characterize the glove film’s mechanical behavior. D6319 addresses these properties before and after aging because storage and time can affect elastomer performance.
Do not publish or contract around a number without its edition, units, conditioning, specimen preparation, test method, and pass rule. A report should show both required and measured results, not simply “pass.” Buyers should also keep shelf-life and storage evidence separate from an accelerated-aging test result.
Residual powder
ASTM D6124 is a test method for residual powder on medical gloves. It reports retained mass under the defined procedure; it is not a general cleanliness, sensitization, extractables, or biocompatibility test.
In the US, powdered patient-examination and surgical gloves are prohibited. A “powder-free” claim and a D6124 result therefore need to be understood inside the current regulatory and product-labeling context. Buyers should not confuse low measured residue with proof that a glove contains no process chemicals or cannot cause irritation.

How AQL works—and what it does not mean
AQL, often expanded as acceptance quality limit, belongs to an attribute-sampling plan. The plan defines a lot, inspection level, sample size, and acceptance/rejection numbers. Inspectors test a sample and make a lot-disposition decision under that plan.
An AQL value is not a promise that exactly that percentage of every box is defective. It is not a probability that one randomly selected glove has a hole. It is also not meaningful when detached from the inspection level, code letter or sample size, and acceptance/rejection criteria.
Some products or contracts use a stricter AQL than the base requirement. That stricter result is product- and evidence-specific; it must not be rewritten as the universal ASTM D6319 threshold. When comparing suppliers, request:
- the standard and edition;
- the sampling standard and inspection level;
- lot definition and sample size;
- acceptance and rejection numbers;
- defect classification;
- actual result and lot disposition;
- laboratory or quality-unit identity and test date.
This detail also prevents a marketing shortcut: “AQL tested” without a number, plan, property, or report is incomplete.
ASTM D6319 versus D6978
ASTM D6978 is not a higher general grade of D6319. The active D6978-05(2023) practice assesses resistance of medical-glove material to permeation by nine chemotherapy drugs under continuous-contact conditions intended to approximate a worst-case clinical scenario. ASTM says the data are for relative comparison of glove materials.
A D6978 result does not prove resistance to every chemotherapy or hazardous drug, every concentration, every duration, or every real-use condition. It also does not establish broad resistance to cleaners, solvents, fuels, or industrial mixtures. The buyer needs the exact tested glove, drug list, concentrations, breakthrough results, test conditions, and use instructions. Applicable occupational and healthcare procedures still control glove changes and other PPE.
INTCO’s Advancare nitrile examination-glove page states that models are tested to ASTM D6978. Treat that as a starting claim: request the exact model’s current report and drug-by-drug data before assigning it to a hazardous-drug task. Avoid the phrase “ASTM D6978 certified” unless a defined certification scheme and certificate support it; ASTM D6978 itself is a test practice.
Where EN 455 and ISO 11193 fit
Buyers sourcing globally may encounter EN 455 and ISO 11193 alongside ASTM D6319. These are not interchangeable labels.
- EN 455 is a multi-part European standard series for single-use medical gloves, covering areas such as freedom from holes, physical properties, biological evaluation, and shelf-life determination. European conformity also depends on the applicable legislation, current harmonized references, and the manufacturer’s conformity route.
- ISO 11193-1 addresses single-use medical examination gloves made from rubber latex or rubber solution. The current published edition and any revision status should be checked with ISO before use.
- EN ISO 374 concerns chemical and microorganism risks for protective gloves. A chemical-protection marking does not replace D6319 medical-examination requirements, and D6319 does not replace claim-specific chemical testing.
The European Commission’s harmonized-standards page explains that voluntary use of standards whose references are published in the Official Journal can provide presumption of conformity for the requirements they cover. Buyers should verify the actual cited edition and legal status rather than copying a supplier icon.
What ASTM D6319 does not prove
D6319 alone does not demonstrate:
- suitability for surgery or a sterile procedure;
- protection against every virus, microorganism, or bloodborne-pathogen exposure scenario;
- puncture resistance to needles or sharps;
- chemotherapy-drug or general chemical permeation resistance;
- food-contact status, cleanroom cleanliness, or ESD performance;
- freedom from all accelerators, irritants, or allergens;
- fit for every user, task, duration, or workplace;
- compliance in every country or region;
- consistent performance of a different SKU in the same product family.
Testing is evidence for a defined claim under defined conditions. It is not permission to expand the claim beyond its scope.
How to review a glove test report
Begin at the top of the report and verify identity before reading the result:
- Applicant and manufacturer: confirm legal names, factory, and addresses.
- Product identity: match brand, model, catalog number, material, color, size, thickness, and packaging.
- Sample identity: check lot, manufacture date, receipt date, condition, and who selected the sample.
- Method: confirm the standard, edition, deviations, conditioning, and referenced equipment.
- Sampling: identify sample size, inspection level, AQL, and acceptance/rejection rule where applicable.
- Results: compare measured values with the correct criterion and units; investigate omissions or unexplained averages.
- Laboratory: verify accreditation scope where claimed, report authorization, location, and issue date.
- Claim match: ensure the report supports the exact intended use and marketing wording.
- Change control: ask what triggers retesting—formulation, pigment, former, line, factory, packaging, or standard revision.

Do not accept a cropped first page, an undated logo, or a report for “representative nitrile gloves” without an identity bridge to the purchased product. If multiple sizes or colors are grouped, request the manufacturer’s rationale and confirm that the test and market file allow that grouping.
Applying the checklist to INTCO glove pages
The INTCO nitrile glove category is useful for finding candidate families, but category breadth also demonstrates why no single certification statement can cover every nitrile product and use.
For medical sourcing, the disposable nitrile examination-glove family lists D6319-related physical information and multiple configurations. A procurement file should still name the exact model, sizes, market, labeling, current reports, and production source.
For process context, INTCO’s article on how disposable-glove manufacturers test quality explains factory and laboratory workflow. That page addresses how testing is performed in general; this article addresses D6319 and how buyers should interpret the evidence.
Buyer questions for a supplier
- Which exact SKU and intended use does this report support?
- Which standard edition and market requirement were applied?
- Is D6319 the product specification, or are you showing only one referenced method?
- How was the lot sampled, and what were the AQL, inspection level, sample size, and Ac/Re numbers?
- Are before- and after-aging physical results included?
- Does “powder-free” match US medical-glove requirements and the product label?
- Which additional claims—chemical, chemotherapy, viral, sterile, food contact, cleanroom, or ESD—have separate evidence?
- What changes require revalidation or customer notification?
- Can the report be traced to the purchased size, color, lot, factory, and package?
FAQ
Is ASTM D6319 a test or a specification?
It is a specification for nitrile examination gloves for medical application. It contains or references procedures and requirements; individual properties can also use separate ASTM methods.
Is ASTM D5151 a puncture test?
No. It is a water-leak method for detecting holes in medical gloves. Sharps and puncture hazards require separate evaluation and controls.
Does AQL 1.5 mean only 1.5% of gloves are defective?
No. AQL belongs to a defined sampling and lot-acceptance system. It is not a literal defect percentage guaranteed for every box or lot.
Does ASTM D6978 mean a glove protects against all chemotherapy drugs?
No. Review the exact glove, tested drugs, conditions, and results. D6978 is a specific permeation assessment, not a universal chemical-resistance approval.
Can a buyer rely on a product-page certification icon?
Not by itself. Ask for the current underlying report and match it to the exact SKU, standard edition, intended use, market, and claim.
The strongest purchasing statement is therefore not “ASTM tested.” It is: this exact finished SKU, in the specified configuration, was evaluated to the named current method under the documented sampling plan and produced results that support this defined claim.